Frequency bandwidth: DC to 100 MHz
- DETAILS
-
- FEATURES
Applicable for all Digital Oscilloscopes or ScopeCorders.
With increases in the speed of measured signals in recent years, it has become necessary to make floating measurements over a wide frequency band. The 700924 differential probe lets you make wide-band (100 MHz) differential input measurements by just connecting the probe to the input of a single-end input digital oscilloscope.
Oscilloscope input amplifiers are normally single-ended. They do not provide accurate measurements when used in combination with a 10:1 passive probe for measuring the gate-to-source voltage of a switching power supply, or potential difference with a common mode voltage in elements such as Insulated Gate Bipolar Transistors (IGBTs).
Furthermore, there is a risk that current will enter through the oscilloscope ground line and damage the oscilloscope.
- SPECIFICATIONS
Input type Balanced differential input Frequency band DC up to 100 MHz (-3 dB) Input attenuation ratio Switched between 1/100 and 1/1000 Input impedance Approximately 4 MΩ, approximately 10 pF parallel Gain accuracy ±2% (common mode voltage: 400 V or less)
±3% (common mode voltage: 1000 V or less)Maximum allowed
differential voltage±350 V (DC + ACpeak) or 250 Vrms (1/100 range)
±1400 V (DC + ACpeak) or 1000 Vrms (1/1000 range)Maximum common mode
input voltage±1400 V (DC + ACpeak) or 1000 Vrms (both 1/100 and 1/1000 ranges) Maximum input voltage ±1400 V (DC + ACpeak) or 1000 Vrms (1/1000 range) Common Mode Rejection Ratio (CMRR) -80 dB (60 Hz), -50 dB (1 MHz) Output voltage ±3.5 V Output impedance Used with 1 MΩ input impedance Power supplies Dry cells, four R6P (SUM-3); when using the AC/DC adapter, use an output voltage of 6 V/200 mA or greater/center plus Dry cell life Approximately 2 hours under continuous use External dimensions (approximate) 207 mm x 83 mm x 38 mm¹ Total length Approximately 1.5 meters Weight Approximately 800 grams (excluding dry cells) Approved standards EMC: EN55011: 1991 + A1: 1997 + A2: 1996
EN50082-1: 1992
Safety: EN61010-1: 1993 + A2: 1995
EN61010-2-031: 1994¹: 38 mm is listed as the maximum value.
Note: Dry cells and AC/DC adapter not included.Model and Code
Model Name Code Description Differential probe 700924 DC to 100 MHz band Each 700924 differential probe comes standard with red and black pincher tips (one each).
- FEATURES